Detil Produk Sound Level Meter Model 831C Larson Davis

Sound Level Meter Model 831C Larson DavisStandards Met by Model 831CThe Model 831C meets the specifications of the following standards :Sound Level Meter StandardsIEC61672-1 Ed. 2.0 ( 2013 ) Class 1, Group XIEC60651 Ed 1.2 ( 2001 ) plus Amendment 1 (1993-02 ) and Amendment 2 ( 2000-10 ) Type 1, Grou
Dapatkan penawaran terbaik untuk Sound Level Meter Model 831C Larson Davis !
Nama
Jumlah Order
Unit
IndoTrading.com adalah media iklan khusus perusahaan, distributor, Grosir, Agen, Supplier, Trading dan Pabrik. Jika anda tertarik untuk membeli Sound Level Meter Model 831C Larson Davis ini dengan harga murah, Mohon hubungi Kontak Perusahaan ini diatas.

Jika anda adalah perusahaan yang ingin menjual di IndoTrading.com ,
Detil Produk

Sound Level Meter Model 831C Larson Davis

Standards Met by Model 831C

The Model 831C meets the specifications of the following standards :

Sound Level Meter Standards

IEC61672-1 Ed. 2.0 ( 2013 ) Class 1, Group X

IEC60651 Ed 1.2 ( 2001 ) plus Amendment 1 (1993-02 ) and Amendment 2 ( 2000-10 ) Type 1, Group X

IEC60804 ( 2000-10 ) Type 1, Group X

ANSI S1.4-2014 Class 1

ANSI S1.43-1997 Type 1

DIN 45657

Octave Filter Standards ( Option 831C- OB3 )

IEC61260 Ed. 2.0 ( 2014 ) Class 1, all filters

ANSI S1.11-2014 Class 1, all filters

Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use

2006/95/EC Low Voltage Safety Directive

IEC 61010-1 Ed. 3.0 ( 2010-06 )

EMC Immunity and Emission

2004/08/EC EMC Directive

IEC 61326-1 Ed. 2.0 ( 2012-07 )

IEC 61672-1 Ed. 2.0 ( 2013-09 )

FCC Title 47 CRF Part 15, Class B

Sound Level Meter Specifications

Averaging ( Integration Method )

Linear or Exponential

Time Weightings

Slow, Fast,or Impulse

Frequency Weightings

A, C, and Z

Peak Detector Frequency Weighting

A, C, or Z

Gain

0 dB or +20 dB

Sample Rate

51,200 Hz

Peak Rise Time

30 μs

Metrics Measured

Leq, Lmax, Lmin, Lpeak, Ln ( 6 values ), LDN, LDEN,LCeq–LAeq

Physical Characteristics

Length with Microphone and Preamplifier

11.35 in

29.0 cm

Length, Instrument Body Only

8.8 in

22.4 cm

Width

2.8 in

7.1 cm

Depth

1.6 in

4.1 cm

Weight with Batteries, No Preamplifer or Microphone

17.3 oz

490 g

General Specifications

Reference Level

114.0 dB re. 20 μPa

Reference Level Range

Single large range for SLM measurements

Reference Frequency

1000 Hz

Reference Direction

0° is perpendicular to the microphone diaphragm

Operating Temperature

≤± 0.5 dB error between -22 °F to +122 °F ( -30 °C to 50 °C )

Storage Temperature

-40 °F to 176 °F ( -40 °C to 80 °C )

Humidity

≤± 0.5 dB error from 30 % to 90 % relative humidity at 104 °F ( 40 °C )

Equivalent Microphone Impedance

12 pF

Effect of an Extension Cable

None up to 200 ft ( 61 m ) with EXCxxx cable

Approvals

CE, ROHS, WEEE

Extended Weather Options

-40 °F to +158 ºF ( -40 °C to +70 ºC ) operation with CER-831-E

Resolution Specifications

Levels

0.1 dB

Elapsed Time

0.1 s

Real Time Clock

1 s

Integration Time

Time Averaged Levels and Sound Exposure Levels

Minimum

0.1 s

Maximum with Daily Autostore Enabled

Unlimited

Maximum with Daily Autostore Disabled

>23 days with error <0.5 dB

Ln Statistics

Number of Selectable Parameters 6 in xx.xx% format

Distribution Resolution 0.1 dB

Spectral Statistics Requires Octave Analysis option ( 831C-OB3 )

Markers

Number of Markers

10

Prenamed Markers

Truck, Automobile, Motorcycle, Aircraft, Exclude

Back Erase

Back Erase Time

5 or 10 s

Measurement Control Modes

Available Modes

Manual Stop, Timed Stop, Stop when Stable, Continuous, Single Block Timer, Daily Block Timer

Timed Stop

Time in hh : mm : ss

Stop When Stable

Delta level in xx.x dB and time in hh:mm:ss

Continuous with Daily Auto-Store

1, 2, 4, 6, 12, 24, 48, 96 or 144 files per day, automated file numbering “yymmddnn.LD0”

Restart after Power Failure

Automatic if powered by 12 VDC and continuous run mode

Single Block Timer

Start date and time to end date and time

Daily Block Timer

Up to 3 blocks between each start and end date

Clock Stability

<1 sec in 24 hours, at 75 ºF ( +24 ºC )

<10 sec in 30 days, at -40 °F to +158 °F ( -40 °C to +70 °C )

<1 s when using NTP