BWB-XP Performance Flame PhotometerTechnical DataSample rate 3.5 – 7 ml/min​Tubing materials Silicone and Tygon®Required desk space for Instrument 50 cm ( H ) x 45 cm ( W ) x 45 cm ( D )*For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of
Dapatkan penawaran terbaik untuk BWB-XP Performance Flame Photometer !
Nama
Jumlah Order
Unit
IndoTrading.com adalah media iklan khusus perusahaan, distributor, Grosir, Agen, Supplier, Trading dan Pabrik. Jika anda tertarik untuk membeli BWB-XP Performance Flame Photometer ini dengan harga murah, Mohon hubungi Kontak Perusahaan ini diatas.

Jika anda adalah perusahaan yang ingin menjual di IndoTrading.com ,
Detil Produk

BWB-XP Performance Flame Photometer

Technical Data

Sample rate 3.5 – 7 ml/min

​Tubing materials Silicone and Tygon®

Required desk space for Instrument 50 cm ( H ) x 45 cm ( W ) x 45 cm ( D )*

For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.

​Instrument size XP Flame Photometer 51 cm ( H ) x 38 cm ( W ) x 41 cm ( D ) ( 20 in x 14 in x 16 in )

​Shipping 62 cm ( H ) x 55 cm ( W ) x 47 cm ( D ) ( 24 in x 22 in x 19 in )

​Weight XP Flame Photometer – 15.3 kg ( 33.7 lbs ) Shipping – 25 kg ( 55.1 lbs )

​Optimal range

Single point calibration

Na 0.1 – 60 ppm

K 0.05 – 100 ppm

Li 0.05 – 50 ppm

Ca 1.0 – 100 ppm

Ba 1.6 – 100 ppm

Multi Point Calibration

Na 0.1 – 1000 ppm

K 0.05 – 1000 ppm

Li 0.05 – 1000 ppm

Ca 1.0 – 1000 ppm

Ba 5.0 – 3000 ppm

Reproducibility <1 % Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100 ppm or less. ( After instrument stabilisation ).

Specificity Na/K/Li = <0.5 % to each other when equal in concentration at <100 ppm

Limit of detection ( LOD ) and limit of quantification ( LOQ )

LOD LOQ

Na – 0.03 ppm Na – 0.1 ppm

K – 0.02 ppm K – 0.05 ppm

Li – 0.02 ppm Li – 0.05 ppm

Ca – 0.3 ppm Ca – 1 ppm

Ba – 1.6 ppm Ba – 5.0 ppm

​Time to stability Less than 15 seconds after sample is introduced into the flame.

​Ions measured Simultaneous measurement of Na, K, Li, Ca, Ba

Interfaces

USB

0-2.5 volt output ( based on sample concentration linked to element of users choice )

Optional 4-20 mA output in place of the above

Optional integrated printer

csv and .pdf generated reports and files via FP-PC software

Recommended minimum warm up time based on ambient temperature

21 ºC – 40 minutes

Power requirements

100V – 250V AC at 50 or 60Hz automatically selecting

Fuel requirement

Propane, Butane or Natural Gas* regulated to 19 Bar. Flow rate of 0.4 l/min

*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.

Readout

LCD, four line, alpha numeric, back lit.