Detil Produk AA500F Flame Atomic Absorption Spectrometer PG Instruments

AA500F Flame Atomic Absorption Spectrometer PG InstrumentsSpecificationsWavelength Range185 nm- 910 nmMonochromatorCzerny-Turner ConfigurationSpectral Bandwidth0.1 nm, 0.2 nm, 0.4 nm, 1.0 nm, 2.0 nm,( software selectable ).Wavelength Accuracy± 0.15 nmWavelength Reproducibility< 0.05 nmReso
Dapatkan penawaran terbaik untuk AA500F Flame Atomic Absorption Spectrometer PG Instruments !
Nama
Jumlah Order
Unit
IndoTrading.com adalah media iklan khusus perusahaan, distributor, Grosir, Agen, Supplier, Trading dan Pabrik. Jika anda tertarik untuk membeli AA500F Flame Atomic Absorption Spectrometer PG Instruments ini dengan harga murah, Mohon hubungi Kontak Perusahaan ini diatas.

Jika anda adalah perusahaan yang ingin menjual di IndoTrading.com ,
Detil Produk

AA500F Flame Atomic Absorption Spectrometer PG Instruments

Specifications

Wavelength Range

185 nm- 910 nm

Monochromator

Czerny-Turner Configuration

Spectral Bandwidth

0.1 nm, 0.2 nm, 0.4 nm, 1.0 nm, 2.0 nm,( software selectable ).

Wavelength Accuracy

± 0.15 nm

Wavelength Reproducibility

< 0.05 nm

Resolution

0.2 nm ± 0.02 nm

Baseline Stability

0.005 A/30 min

Sensitivity ( Cu )

2 µg/ml Absorption > 0.28A ( flame )

Detection Limit

Cu < 0.004 µg/ml ( flame ) Cd < 0.4 x 10-12 g ( graphite furnace )

Repeatability

Cu < 0.7 % ( Air/Acetylene flame )

Ba < 1.0 % ( Nitrous oxide/Acetylene flame ) Cu < 2.0 % Cd < 2.0 % ( Graphite Furnace )

Background Correction

Deuterium Arc, Self reversal

Characteristic Concentration

Cu < 0.02 µg/ml, Ba < 0.15 mg/ml ( N₂O/Acetylene )

Burner Heads

Titanium Alloy

Nebuliser

Inert Adjustable Nebuliser, Flow Rate from 2 ml – 6 ml/min. Pt/Ir capillary available for high acid concentration

Atomization Chamber

Corrosion-resistant material

Position Adjustment

Automatic changeover ( AA500GF ) Manual ( AA500F ) Automatic Settingof Optimum Height for Flame Burner.

Safety Functions

Burner Identification, Flame Sensor, Gas leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water ( graphite ), Over Temperature Sensor ( graphite )